The iodine sorption-filtering tape SFL-2I-50 is made of a material FPP with high disperse sorbent and intended for culling hallmarks of aerosols, both vapours of iodine and subsequent measuring of the delayed sediment by radiometric instruments.
The tape SFL-2I-50 represents 3-beds band composed of a friable stratum of ultra thin fibres, into which one is entered impregnated by coal covered from both parties with strata of a tighter material, such as NEL. Edges of the tape are pressure-tight, fastened by calorific welding.
The tape SFL-2I-50 is exhausted in a communal width of 50 mm. Width of a working part without the fastened edges of 40 mm; permissible variation is ±1 mm.
The tape is coiled onto a rigid core with an inside diameter of 20 mm. Length of a tape reeled onto the core is 25-35 m.
The tape SFL-2I-50 can be used as a semi-finished item for fabrication of analytical screens AFAS-I, also contents, intended for definition, radioactive iodine.
The filtering ground (basis) of a tape is hydrophobic, the rack to strong acids, and also alcohols, is diluted in chlorinated hydrocarbons and ketones.
CHARACTERISTICS
| The naming of indexes | N o r m | ||
| Group 1 | Group 2 | ||
| 1 | Resistance of a tape to an air stream at a speed of 1cm/s, Pа | 25-55 | 55-75 |
| 2 | Penetrations coefficient on an oily fog with a particle radius of 0.15-0.17 ?m at a filtration rate of 10 cm/s, %, no more | 1,0 | 1,0 |
| 3 | Dynamic capacitance on molecular stable iodine at a speed of 1 cm/s, concentration 2.5 mg/l and penetrations coefficient of iodine of 0.001 %, mg/cm2, not less | 0,5 | 0,5 |
| 4 | Penetrations coefficient molecular radioactive iodine - 131 at a speed of 5 l/min, initial concentration 1.10-7 –1.10-8 Curie/l, floor space of a sample 3 см2, temperature 18-22°С and iodine–131 capacitance 1.10-5 Curie, %, no more | 1,0 | 1,0 |
| 5 | Mass of a unit area of a tape, mg/cm2 | 10-20 | 10-20 |
| 6 | Breaking strength of a tape with the reinforced edges, N, not less | 50 | 50 |
The tape SFL-2I-50 is applied in radiometric instruments to catching and defining the contents of radioactive iodine, counted on exploitation at temperature of environmental and monitored air from 0?С up to +60?С with relative humidity up to 100 % (without formation of a dropped moisture) and the stay (is absent hesitation pumping through a tape of monitored air) at temperature from -50?С up to +65?С at the same limiting value.
The conformity of tapes to the technical requirements within 5 years from the moment of fabrication is guaranteed at observance of conditions of storage and transportation.